Laboratory objectives
What you will investigate
By the end of this virtual laboratory, you should be able to connect cardiac electrical activity, ECG waveform features and lead mathematics to the signals acquired by a biomedical instrument.
Experiment 1
Virtual ECG waveform and heart-rate simulator
- Relate P, QRS and T waves to atrial and ventricular activity.
- Change waveform amplitudes and calculate rate from the R–R interval.
- Observe the effect of noise and calibration settings.
Experiment 2
Virtual 3-lead placement and Einthoven’s law
- Correctly place RA, LA and LL electrodes.
- Define leads I, II and III as voltage differences.
- Verify that II = I + III experimentally.
Experiment 3
Augmented unipolar lead system: aVR, aVL, aVF
- Derive augmented leads from the Wilson central terminal.
- Compare aVR, aVL and aVF polarities with the cardiac vector.
- Explain the identity aVR + aVL + aVF = 0.
Background theory
From cardiac activity to limb-lead voltages
ECG waves and rate
The P wave represents atrial depolarisation, the QRS complex represents ventricular depolarisation and the T wave represents ventricular repolarisation. In a regular rhythm, the heart rate is calculated from the R–R interval.
At 25 mm/s: 1 small square = 0.04 s, 1 large square = 0.20 s
HR = 300 / number of large squares between R waves
Standard limb electrodes
For limb leads, the right arm (RA), left arm (LA) and left leg (LL) electrodes form three bipolar leads. A right-leg electrode may provide a reference or driven common-mode connection.
Lead II = VLL − VRA
Lead III = VLL − VLA
Augmented unipolar limb leads
The Wilson central terminal averages the three limb potentials. A unipolar limb lead is the selected limb potential minus the central-terminal potential. “Augmentation” multiplies the signal by 3/2 so that the resulting aVR, aVL and aVF have more practical amplitude.
aVR = VRA − ½(VLA + VLL)
aVL = VLA − ½(VRA + VLL)
aVF = VLL − ½(VRA + VLA)
Learning relationship
Adding the three standard limb-lead equations gives:
Axis convention
Lead I is 0°, lead II +60°, lead III +120°, aVL −30°, aVF +90° and aVR −150° in the frontal plane.
Clinical significance
aVR, aVL and aVF provide frontal-plane views; changes in their polarity and amplitude help infer electrical axis and regional patterns.
Virtual equipment and conventions
Laboratory setup
Virtual equipment
- ECG waveform generator
- RA, LA, LL and reference electrode models
- Lead I, II and III voltmeters
- Augmented-lead computation module
- Digital oscilloscope / ECG display
- Data recorder and CSV export
Measurement conventions
- Amplitudes are displayed in millivolts (mV).
- Time is displayed in seconds (s); standard ECG paper speed is 25 mm/s.
- The cardiac vector model is educational and simplified.
- RA is assigned 0 V in some calculations only as a relative reference; all lead conclusions remain unchanged.
Experiment 1 · Waveform and rate
Virtual ECG waveform and heart-rate simulator
Vary the heart rate and P/QRS/T amplitudes, observe a simulated ECG trace and calculate heart rate from the R–R interval. Add noise to examine the engineering challenge of obtaining a clean diagnostic signal.
75 bpm
0.80 s
4.00
12.5
Procedure
- Set the heart rate to 75 bpm and record the theoretical R–R interval.
- Reduce and increase heart rate; state how the waveform spacing changes.
- Change P, QRS and T amplitudes independently and identify each feature on the display.
- Add random noise and 50 Hz interference; describe the effect on measurement reliability.
- Record at least six snapshots for different heart rates, then plot calculated against simulated heart rate.
Experiment 1 observations
| No. | HR setting (bpm) | R–R (s) | R–R (large boxes) | Calculated HR (bpm) | Noise (%) | 50 Hz (%) |
|---|
Experiment 2 · Three limb leads
Virtual 3-lead placement and Einthoven’s-law experiment
Drag each electrode to the correct anatomical target. Once the three-limb configuration is valid, inspect the instantaneous voltages and verify that lead II equals lead I plus lead III.
Lead definitions
II = LL − RA
III = LL − LA
Einthoven: II = I + III
0.000 mV
0.000 mV
0.000 mV
0.000 mV
0.000 V
0.000 V
0.000 V
Procedure
- Drag the RA, LA, LL and RL reference electrodes to their anatomical positions.
- Click Check Placement. Correct any red targets and repeat.
- Change the cardiac vector angle and magnitude. Predict which lead should become more positive or negative.
- Move the sample-phase control and verify that II − (I + III) remains zero.
- Record snapshots for at least five vector angles, including values that make one lead approximately zero.
Experiment 2 observations
| No. | Vector angle (°) | Magnitude (mV) | Lead I (mV) | Lead II (mV) | Lead III (mV) | II − (I+III) (mV) |
|---|
Experiment 3 · Augmented unipolar leads
Augmented lead system: aVR, aVL and aVF
Use the cardiac-vector controls to derive the three augmented limb leads, inspect their polarities and verify the identity aVR + aVL + aVF = 0.
Augmented-lead equations
aVL = LA − ½(RA + LL)
aVF = LL − ½(RA + LA)
aVR + aVL + aVF = 0
−0.693 mV
0.000 mV
0.693 mV
0.000 mV
Procedure
- Set the cardiac vector to +60° and record the aVR, aVL and aVF amplitudes.
- Rotate the vector toward +90° and explain the change in aVF and aVL.
- Find the angle that makes aVL approximately zero and the angle that makes aVF approximately zero.
- Verify that the algebraic sum of the three augmented leads remains zero.
- Predict what happens when the vector points directly toward aVR.
Experiment 3 observations
| No. | Vector angle (°) | Magnitude (mV) | aVR (mV) | aVL (mV) | aVF (mV) | Sum (mV) |
|---|
Laboratory report
Guidelines for report writing
Required structure
- Title page: experiment title, course code, student name, registration number, group and date.
- Objectives: state the purpose of each of the three experiments.
- Theory: summarise ECG waves, Einthoven’s law and augmented-lead equations.
- Equipment and setup: describe the virtual modules and control settings.
- Procedure: write the steps in your own words and include screenshots where useful.
- Results: present observation tables, calculations and waveform screenshots.
- Discussion: explain trends, compare expected and measured results, and identify likely sources of discrepancy.
- Conclusion: answer each objective in one or two concise statements.
- References: cite textbooks, standards and course materials used.
Discussion prompts
- How does the R–R interval change when heart rate doubles?
- Why is lead II equal to lead I plus lead III?
- What happens to aVR, aVL and aVF when the cardiac vector is +90°?
- Why can electrode misplacement change the calculated cardiac axis?
- Which filtering approach would reduce 50 Hz interference without removing clinically useful QRS detail?
Submission checklist
- All three procedures completed
- At least six Experiment 1 records
- At least five Experiment 2 records
- At least five Experiment 3 records
- CSV export attached or data tables reproduced
- Calculations and units included