Introduction to Biomedical Engineering — Undergraduate ECE
This virtual laboratory lets you explore a multi-parameter bedside patient monitor of the kind found in every Intensive Care Unit (ICU). You will observe real-time physiological waveforms (ECG, SpO₂ plethysmograph, respiration), read numeric vital-sign tiles, configure alarm limits, and respond to simulated clinical events such as arrhythmia, hypoxia, hypotension, ventricular tachycardia, and asystole — all from your browser, with no equipment required.
After completing this experiment, you should be able to:
A bedside patient monitor is a real-time biomedical instrumentation system. The sections below summarise the theory needed for this experiment.
A patient monitor continuously acquires weak physiological signals from the body, amplifies and conditions them, converts them to digital form, and presents waveforms, numeric values, and alarms to the clinical staff. The general acquisition chain is:
The front-end amplifier is typically an instrumentation amplifier with high common-mode rejection ratio (CMRR > 100 dB), because biopotentials such as the ECG are only 0.5–4 mV while interference can be volts. Isolation barriers (optical or magnetic) protect the patient from leakage currents. The ADC typically samples at 250–1000 Hz per channel, and the microprocessor extracts features (heart rate, SpO₂, arrhythmias) in real time.
| Parameter | Sensor / Method | Unit | Normal Range (adult) |
|---|---|---|---|
| ECG / Heart Rate | 3-lead or 5-lead Ag/AgCl electrodes; Lead II is standard for rhythm monitoring | bpm | 60–100 bpm |
| SpO₂ (oxygen saturation) | Pulse oximetry — Beer–Lambert law; two LEDs: red 660 nm and infrared 940 nm | % | 95–100 % |
| NIBP (non-invasive blood pressure) | Oscillometric cuff method (clinically validated against Korotkoff sounds) | mmHg | ≈120/80 mmHg |
| Respiration rate | Impedance pneumography via ECG electrodes, or capnography | breaths/min | 12–20 /min |
| Temperature | Thermistor skin/esophageal probe (negative temperature coefficient) | °C | 36.5–37.5 °C |
| IBP (optional) | Invasive arterial line with pressure transducer; shows beat-by-beat arterial waveform | mmHg | ≈120/80 mmHg |
Pulse oximetry relies on the different absorption of oxygenated and de-oxygenated haemoglobin at two wavelengths (660 nm red, 940 nm infrared). The ratio of the pulsatile (AC) to steady (DC) absorbance at each wavelength is combined into a ratio-of-ratios R, empirically mapped to SpO₂.
R = (AC660/DC660) / (AC940/DC940) → SpO₂ ≈ 110 − 25·RMean Arterial Pressure (MAP) is estimated from the oscillometric NIBP reading as:
MAP = (SYS + 2·DIA) / 3For a reading of 120/80 mmHg, MAP = (120 + 2×80)/3 ≈ 93 mmHg.
The electrocardiogram records the heart's electrical activity from the body surface. One cardiac cycle in Lead II consists of the PQRST complex:
Heart rate is computed from the interval between consecutive R peaks (the RR interval):
HR (bpm) = 60 / RR-interval (seconds)Example: an RR interval of 0.83 s corresponds to HR ≈ 72 bpm. At standard 25 mm/s sweep, one small (1 mm) division = 0.04 s and one large (5 mm) division = 0.20 s, so HR can also be estimated as 300 / (number of large divisions between R peaks).
Monitors compare each vital sign against configurable high and low limit alarms. The IEC 60601-1-8 standard defines alarm priority levels, colours, and audible patterns:
Excessive or false alarms cause alarm fatigue: clinicians become desensitised and may miss genuine events. Properly setting patient-specific alarm limits and using the silence function correctly are essential safety skills you will practise in this lab.
A typical monitoring-mode ECG bandwidth is 0.5–40 Hz (a bandpass filter), which suppresses drift and high-frequency noise while preserving the QRS shape for rhythm analysis. Diagnostic ECG uses a wider 0.05–150 Hz bandwidth. In the simulation you can add noise and toggle a 50 Hz notch filter to see its effect directly.
Because electrodes create a low-impedance path to the body (and catheters can reach the heart directly), patient-connected equipment must meet IEC 60601-1 leakage-current limits. Patient leakage current must stay in the microampere range (as low as 10 µA for cardiac-applied parts), because currents as small as tens of µA delivered directly to the heart can induce ventricular fibrillation (microshock). This is achieved with isolated front-ends, isolated power supplies, and protective earthing — the “isolation” block in the diagram of Section 4.1.
A fully interactive patient monitor. Select a clinical scenario, watch the waveforms and vital signs change, manage alarms, and experiment with noise and filtering. Click any control to enable alarm sound (your browser requires a click before audio can play).
Perform the steps in order and record all results in the Observations table. Take a screenshot of the monitor at each step and include it in your report.
Fill in the table during the experiment. Default alarm limits: HR 50–120 bpm, SpO₂ ≥ 90 %, SYS 90–160 mmHg, RESP 8–30 /min.
| Scenario | HR (bpm) | SpO₂ (%) | NIBP (mmHg) | RR (/min) | Alarms Triggered | Student Notes |
|---|---|---|---|---|---|---|
| Normal Sinus Rhythm | ||||||
| Sinus Tachycardia | ||||||
| Bradycardia | ||||||
| Hypoxia | ||||||
| Hypotension | ||||||
| Ventricular Tachycardia | ||||||
| Asystole | ||||||
| Motion Artifact |
| Condition | ECG appearance | Alarms | Student Notes |
|---|---|---|---|
| Noise = 0, Notch OFF | |||
| Noise = 8, Notch OFF | |||
| Noise = 8, Notch ON |
Your lab report should follow this structure: