Biomedical Virtual Labs

Virtual Laboratory: Patient Monitor / ICU Dashboard Simulation

Introduction to Biomedical Engineering — Undergraduate ECE

This virtual laboratory lets you explore a multi-parameter bedside patient monitor of the kind found in every Intensive Care Unit (ICU). You will observe real-time physiological waveforms (ECG, SpO₂ plethysmograph, respiration), read numeric vital-sign tiles, configure alarm limits, and respond to simulated clinical events such as arrhythmia, hypoxia, hypotension, ventricular tachycardia, and asystole — all from your browser, with no equipment required.

Objectives

After completing this experiment, you should be able to:

  • Describe the architecture of a multi-parameter patient monitor, from electrode/sensor front-end through amplification, filtering, isolation, A/D conversion, and display.
  • Identify and interpret the standard monitored signals: ECG (Lead II) and heart rate, SpO₂ plethysmograph, respiration waveform, non-invasive blood pressure (NIBP), and body temperature.
  • Configure high/low alarm limits for vital parameters and respond appropriately to visual and audible alarm events of different priority levels.
  • Observe and document the physiological effect of clinical scenarios: sinus tachycardia, bradycardia, hypoxia, hypotension, ventricular tachycardia, and asystole.
  • Recognise signal artifacts (power-line interference, baseline wander, motion artifact) and evaluate the effect of a 50 Hz notch filter on the ECG trace.
  • Relate monitor behaviour to patient-safety standards (IEC 60601-1, IEC 60601-1-8) and the clinical problem of alarm fatigue.

Theory

A bedside patient monitor is a real-time biomedical instrumentation system. The sections below summarise the theory needed for this experiment.

4.1 What is a Patient Monitor?

A patient monitor continuously acquires weak physiological signals from the body, amplifies and conditions them, converts them to digital form, and presents waveforms, numeric values, and alarms to the clinical staff. The general acquisition chain is:

Patient
➞
Electrodes /
Sensors
➞
Front-end
Amplifier
➞
Filtering &
Isolation
➞
ADC
➞
Micro-
processor
➞
Display + Alarms
+ Network

The front-end amplifier is typically an instrumentation amplifier with high common-mode rejection ratio (CMRR > 100 dB), because biopotentials such as the ECG are only 0.5–4 mV while interference can be volts. Isolation barriers (optical or magnetic) protect the patient from leakage currents. The ADC typically samples at 250–1000 Hz per channel, and the microprocessor extracts features (heart rate, SpO₂, arrhythmias) in real time.

4.2 Measured Parameters

ParameterSensor / MethodUnitNormal Range (adult)
ECG / Heart Rate3-lead or 5-lead Ag/AgCl electrodes; Lead II is standard for rhythm monitoringbpm60–100 bpm
SpO₂ (oxygen saturation)Pulse oximetry — Beer–Lambert law; two LEDs: red 660 nm and infrared 940 nm%95–100 %
NIBP (non-invasive blood pressure)Oscillometric cuff method (clinically validated against Korotkoff sounds)mmHg≈120/80 mmHg
Respiration rateImpedance pneumography via ECG electrodes, or capnographybreaths/min12–20 /min
TemperatureThermistor skin/esophageal probe (negative temperature coefficient)°C36.5–37.5 °C
IBP (optional)Invasive arterial line with pressure transducer; shows beat-by-beat arterial waveformmmHg≈120/80 mmHg

Pulse oximetry relies on the different absorption of oxygenated and de-oxygenated haemoglobin at two wavelengths (660 nm red, 940 nm infrared). The ratio of the pulsatile (AC) to steady (DC) absorbance at each wavelength is combined into a ratio-of-ratios R, empirically mapped to SpO₂.

R = (AC660/DC660) / (AC940/DC940)   →   SpO₂ ≈ 110 − 25·R

Mean Arterial Pressure (MAP) is estimated from the oscillometric NIBP reading as:

MAP = (SYS + 2·DIA) / 3

For a reading of 120/80 mmHg, MAP = (120 + 2×80)/3 ≈ 93 mmHg.

4.3 The ECG Waveform

The electrocardiogram records the heart's electrical activity from the body surface. One cardiac cycle in Lead II consists of the PQRST complex:

  • P wave — atrial depolarisation (≈0.08–0.10 s, small rounded upward wave).
  • QRS complex — ventricular depolarisation (≈0.06–0.10 s; sharp large spike, up to 1–4 mV). A wide or bizarre QRS suggests a ventricular origin, as in ventricular tachycardia.
  • T wave — ventricular repolarisation (rounded wave following the QRS).
  • PR interval (0.12–0.20 s) and QT interval (rate-dependent, ≈0.35–0.44 s) are important diagnostic intervals.

Heart rate is computed from the interval between consecutive R peaks (the RR interval):

HR (bpm) = 60 / RR-interval (seconds)

Example: an RR interval of 0.83 s corresponds to HR ≈ 72 bpm. At standard 25 mm/s sweep, one small (1 mm) division = 0.04 s and one large (5 mm) division = 0.20 s, so HR can also be estimated as 300 / (number of large divisions between R peaks).

4.4 Alarm Systems

Monitors compare each vital sign against configurable high and low limit alarms. The IEC 60601-1-8 standard defines alarm priority levels, colours, and audible patterns:

  • High priority (red, fast repeating tone) — immediate threat to life, e.g. asystole, ventricular tachycardia, severe desaturation. Requires immediate response.
  • Medium priority (yellow/amber, slower tone) — abnormal condition needing prompt attention, e.g. HR above limit.
  • Low priority / advisory (cyan/white) — technical or informational alerts, e.g. lead off.

Excessive or false alarms cause alarm fatigue: clinicians become desensitised and may miss genuine events. Properly setting patient-specific alarm limits and using the silence function correctly are essential safety skills you will practise in this lab.

4.5 Artifacts & Filtering

  • Power-line interference (50/60 Hz) — capacitive coupling from mains wiring adds a sinusoidal hum. Removed by a narrow notch filter at 50 Hz (or 60 Hz).
  • Baseline wander — slow drift (<0.5 Hz) caused by respiration and electrode impedance changes; removed by a high-pass filter.
  • Motion artifact / EMG noise — patient movement and muscle activity add broadband noise that can mimic arrhythmias and cause false alarms.

A typical monitoring-mode ECG bandwidth is 0.5–40 Hz (a bandpass filter), which suppresses drift and high-frequency noise while preserving the QRS shape for rhythm analysis. Diagnostic ECG uses a wider 0.05–150 Hz bandwidth. In the simulation you can add noise and toggle a 50 Hz notch filter to see its effect directly.

4.6 Electrical Safety

Because electrodes create a low-impedance path to the body (and catheters can reach the heart directly), patient-connected equipment must meet IEC 60601-1 leakage-current limits. Patient leakage current must stay in the microampere range (as low as 10 µA for cardiac-applied parts), because currents as small as tens of µA delivered directly to the heart can induce ventricular fibrillation (microshock). This is achieved with isolated front-ends, isolated power supplies, and protective earthing — the “isolation” block in the diagram of Section 4.1.

Key idea: every design feature of a patient monitor — amplification, filtering, alarms, isolation — exists to extract a weak signal safely and to warn clinicians reliably.

Simulation: ICU Bedside Monitor

A fully interactive patient monitor. Select a clinical scenario, watch the waveforms and vital signs change, manage alarms, and experiment with noise and filtering. Click any control to enable alarm sound (your browser requires a click before audio can play).

ADULT  |  BED 04 — ICU MONITORING • LOCAL --:--:--
ALARM
II  ECG 25 mm/s  |  10 mm/mV
PLETH  SpO₂ 25 mm/s
RESP  IMPEDANCE 25 mm/s
HRII
72bpm
limits 50 – 120
SpO₂%
98%
low limit 90
NIBPmmHg
120/80 (93)
last: --:--:--
RESP/min
14/min
limits 8 – 30
TEMP°C
36.8°C
skin probe
CLINICAL SCENARIOS
MONITOR CONTROLS
0
ALARM LIMITS
HR HIGH (bpm)
HR LOW (bpm)
SpO₂ LOW (%)
SYS HIGH (mmHg)
SYS LOW (mmHg)
RESP HIGH (/min)
RESP LOW (/min)
ALARM LOG
No alarm events yet.
Tip: use Freeze to stop the sweep and measure RR intervals on screen; adjust Noise / artifact and toggle the 50 Hz notch filter to study interference; tighten or widen the alarm limits and observe how alarm frequency changes.

Procedure

Perform the steps in order and record all results in the Observations table. Take a screenshot of the monitor at each step and include it in your report.

  1. Open the simulation and identify each waveform (ECG, plethysmograph, respiration) and each numeric tile (HR, SpO₂, NIBP, RESP, TEMP). Note the colour convention used for each parameter.
  2. Select Normal Sinus Rhythm. Wait for the values to settle, then record the baseline HR, SpO₂, NIBP, and RR in the observations table.
  3. Press Freeze and measure the RR interval on the ECG (use the sweep speed and grid timing). Calculate heart rate using HR = 60 / RR(s) and compare it with the displayed value. Resume the sweep.
  4. Select Sinus Tachycardia, then Bradycardia. For each, record the new HR and note which alarm fires, at which limit, and its priority level (colour and beep pattern).
  5. Induce Hypoxia. Observe the fall in SpO₂ and the compensatory rise in HR and respiration rate. Record the values and the alarms triggered.
  6. Induce Ventricular Tachycardia and then Asystole. Describe the ECG morphology change (wide QRS; flatline), the alarm priority behaviour, and estimate your response time to silence the alarm. Observe what happens to SpO₂ during asystole.
  7. Return to Normal Sinus Rhythm. Increase the Noise / artifact slider step by step. Note the appearance of baseline wander, EMG noise, and 50 Hz hum, and watch for false alarms.
  8. With noise still high, toggle the 50 Hz Notch filter ON. Record the visible difference in the ECG trace and in alarm behaviour.
  9. Adjust the alarm limits (e.g. raise HR-high, lower SpO₂-low) to reduce false alarms. Note the trade-off and relate it to the concept of alarm fatigue.
  10. For three scenarios of your choice, compare your hand-measured HR (from the frozen ECG) with the monitor's displayed HR, and comment on any difference.

Observations

Fill in the table during the experiment. Default alarm limits: HR 50–120 bpm, SpO₂ ≥ 90 %, SYS 90–160 mmHg, RESP 8–30 /min.

ScenarioHR (bpm)SpO₂ (%)NIBP (mmHg) RR (/min)Alarms TriggeredStudent Notes
Normal Sinus Rhythm
Sinus Tachycardia
Bradycardia
Hypoxia
Hypotension
Ventricular Tachycardia
Asystole
Motion Artifact

Filter & Noise Observations

ConditionECG appearanceAlarmsStudent Notes
Noise = 0, Notch OFF
Noise = 8, Notch OFF
Noise = 8, Notch ON

Report Guidelines

Your lab report should follow this structure:

  • Aim — state the purpose of the experiment in 2–3 sentences.
  • Apparatus / Software — the virtual patient monitor simulation (this web page), browser used.
  • Theory — summarise the monitor architecture, measured parameters, ECG waveform, and alarm system in your own words (with the block diagram).
  • Procedure — list the steps you actually performed.
  • Observations & Results — completed observation tables plus screenshots of the monitor for each scenario; include your hand-calculated HR values and MAP calculations.
  • Discussion — interpret the results: compensatory changes during hypoxia, morphology changes in VT/asystole, effect of noise and the notch filter, alarm behaviour and false alarms.
  • Conclusion — what you learned about patient monitoring, signal quality, and alarm management.
  • References — e.g. Webster, Medical Instrumentation; IEC 60601-1 / 60601-1-8 standards; your course notes.

Viva / Discussion Questions

  1. Why is a 50 Hz notch filter needed in ECG monitoring, and what is the risk of making it too wide?
  2. Explain how the oscillometric method estimates systolic, mean, and diastolic pressure from cuff-pressure oscillations. What role do Korotkoff sounds play in the reference (auscultatory) method?
  3. What is alarm fatigue, and how do properly chosen alarm limits and priorities improve patient safety?
  4. Why is galvanic isolation between the patient front-end and the mains-powered parts of a monitor critical? Which leakage-current range does IEC 60601-1 permit?
  5. During hypoxia, why do heart rate and respiration rate rise? Which sensors on the monitor reveal each change?
  6. Why does the pulse oximeter need two wavelengths (660 nm and 940 nm) instead of one?
Academic integrity: record your own observations and screenshots. Identical tables across reports will be treated as copied work.