Instructions
- This test contains 20 multiple-choice questions on echocardiography.
- Questions are limited to the first three levels of Bloom's Taxonomy: Remember (recall facts), Understand (explain concepts), and Apply (use formulas and principles in calculations).
- Select one answer per question, then click Submit Test to get your score and per-question feedback.
- After submitting, review the Post-Test Answer Key & Explanations section at the bottom — read every explanation, including those you got right.
- Closed book. Time suggested: 40 minutes. Total: 20 marks (1 mark each).
Diagnostic cardiac ultrasound typically uses which frequency range?
The average speed of sound in human soft tissue used by ultrasound scanners is approximately:
Echocardiography is preferred over many other cardiac imaging methods primarily because it:
The piezoelectric crystal material traditionally used in ultrasound transducers is:
In A-mode (Amplitude mode) display, the vertical axis represents:
In colour Doppler, the BART convention means:
M-mode echocardiography displays:
Why do the blood-filled chambers of the heart (e.g., the LV cavity) appear black on a B-mode image?
Increasing the transmit frequency improves image resolution but reduces penetration because:
The factor of 2 appears in the Doppler equation fd = 2f₀v·cosθ/c because:
If the ultrasound beam is perpendicular (θ = 90°) to the direction of blood flow, the measured Doppler shift is:
Why does M-mode provide better temporal resolution than 2D B-mode for measuring fast-moving structures?
Aliasing in pulsed-wave Doppler occurs when:
The ALARA principle in diagnostic ultrasound states that exposure should be:
An echo returns 200 µs after transmission. Using c = 1540 m/s, the depth of the reflector is:
A scanner transmits at 3.5 MHz. The theoretical axial resolution (λ/2) is approximately:
With f₀ = 2.5 MHz, blood velocity v = 1.0 m/s and θ = 0°, the Doppler shift (c = 1540 m/s) is:
A PW Doppler system with f₀ = 2.5 MHz measures fd = 1.62 kHz at θ = 0°. The blood velocity is:
Continuous-wave Doppler measures a peak velocity of 4 m/s across the aortic valve. Using the simplified Bernoulli equation (ΔP = 4v²), the peak pressure gradient is:
A phased-array probe with element spacing d = 0.3 mm steers the beam to θ = 30°. The required progressive delay between adjacent elements (c = 1540 m/s) is approximately:
Your Score
Post-Test Answer Key & Explanations
Correct answers are highlighted in green. Study each explanation — the distractors are designed around common student errors.
▸ Show / Hide Full Answer Key
Diagnostic echocardiography uses 2–15 MHz. Below 2 MHz resolution is too poor; above ~15 MHz attenuation prevents adequate penetration of the chest wall. (Human hearing tops out at 20 kHz; 100 MHz+ would be radio frequencies.)
c ≈ 1540 m/s is the standard assumed speed (13 µs round-trip time per cm of depth). 340 m/s is the speed of sound in air — a common distractor.
Echo is non-invasive, portable, repeatable and free of ionising radiation. Resolution is good but not better than MRI, and image acquisition is highly operator-dependent.
PZT — lead zirconate titanate — converts electrical energy to mechanical vibration and vice versa, enabling the same crystal to transmit and receive.
A-mode plots echo amplitude versus depth (range) along a single beam line. It is the raw data from which all other modes are built.
BART: Blue = flow Away from the transducer, Red = flow Towards the transducer. Green/variance is added to encode turbulence.
M-mode (Motion mode) keeps one beam line fixed: depth on the vertical axis, time scrolling horizontally — hence very high temporal resolution.
Echo brightness depends on impedance mismatches. Blood is acoustically homogeneous, so almost no energy scatters back — the cavity is anechoic (black). Walls and valves, with large impedance steps, appear bright.
Attenuation in tissue is roughly proportional to frequency (≈ 0.5–1 dB/cm/MHz round trip). Higher f₀ → shorter wavelength (better resolution, λ/2) but the echo from depth is attenuated more — the classic resolution–penetration trade-off.
The moving red blood cell first 'receives' a Doppler-shifted wave (moving receiver), then re-radiates it as a moving source — producing a double shift, hence the factor 2.
fd ∝ cosθ and cos90° = 0. Flow crossing the beam at right angles is invisible to Doppler — which is why sonographers align the beam as parallel to flow as possible.
2D must spread its pulses over ~100 scan lines per frame (~30–60 frames/s), while M-mode fires repeatedly along ONE line (thousands of lines/s), capturing the exact extremes of motion for precise measurements.
PW Doppler samples at the PRF; the sampling theorem limits unambiguous shifts to ±PRF/2. Beyond that, high velocities 'wrap around' the baseline — the same aliasing you studied in signals & systems.
ALARA — As Low As Reasonably Achievable — directs operators to use the minimum acoustic output that gives a diagnostic image, limiting thermal and mechanical bioeffects.
d = c·t/2 = 1540 × 200×10⁻⁶ / 2 = 0.154 m = 15.4 cm. Remember the factor of 2 for the round trip — 30.8 cm is the classic wrong answer.
λ = c/f₀ = 1540 / 3.5×10⁶ ≈ 0.44 mm; axial resolution ≈ λ/2 ≈ 0.22 mm. Option 0.44 mm forgets the factor of 2; 0.11 mm is the value for 7 MHz.
fd = 2f₀v·cosθ/c = 2 × 2.5×10⁶ × 1.0 × 1 / 1540 ≈ 3247 Hz ≈ 3.25 kHz. (1.62 kHz would result from forgetting the factor of 2.)
v = fd·c/(2f₀cosθ) = 1620 × 1540 / (2 × 2.5×10⁶ × 1) = 0.50 m/s — a normal transmitral flow velocity.
ΔP = 4v² = 4 × 4² = 4 × 16 = 64 mmHg — a severe aortic stenosis gradient (normal aortic valve peak velocity ≈ 1 m/s → ΔP ≈ 4 mmHg).
Δt = d·sinθ/c = 0.3×10⁻³ × sin30° / 1540 = 0.3×10⁻³ × 0.5 / 1540 ≈ 97 ns. These nanosecond delays are generated by the system's digital beamformer.