ECE 331 · Introduction to Biomedical Engineering

Electrocardiography

A comprehensive, interactive study guide for third-year Electrical and Communication Engineering students. It connects cardiac electrophysiology, clinical ECG interpretation, and the electronic instrumentation used to acquire one of medicine’s most important biosignals.

Department of Electrical & Communication Engineering Masinde Muliro University of Science and Technology Level: Third Year Undergraduate
Cardiac electrical signal → engineering measurement → clinical decision

Orientation

What is electrocardiography and why should an engineer study it?

Electrocardiography is the non-invasive recording of the heart’s electrical activity from the body surface. The measured potential differences are only a few millivolts and contain clinically important events lasting milliseconds to seconds. Obtaining a reliable ECG therefore requires electrodes, low-noise analogue amplifiers, interference rejection, filtering, analogue-to-digital conversion, signal processing and clear visualisation.

1

Electrophysiology

Ion movement across cardiac cell membranes creates action potentials that spread through excitable tissue.

2

Electromagnetics & circuits

Body tissues conduct tiny potentials to electrodes. Instrumentation must reject common-mode interference and protect the patient.

3

Signal processing

Filtering and digital processing separate PQRST information from baseline wander, muscle noise and power-line interference.

Learning outcomes

Core idea: the ECG is a time-varying surface map of cardiac electrical vectors, not a direct recording of heart contractions. Electrical activity normally precedes mechanical contraction.

Cellular foundations

Types of heart cells

Cardiac tissue contains specialised cells with different jobs. Some generate rhythmic impulses, some conduct rapidly, and most contract. Their different ion-channel profiles produce distinct action-potential shapes.

Pacemaker cells SA node · AV node · automaticity Conducting cells His bundle · bundle branches · Purkinje fibres Contractile myocytes Atrial and ventricular working muscle
Original educational illustration: pacemaker cells initiate impulses, specialised conducting cells distribute them rapidly, and contractile myocytes generate force.
Cell typeMain locationElectrical behaviourPrimary functionECG significance
SA nodal pacemaker cellsRight atrium near the superior vena cavaSpontaneous phase-4 depolarisation; slow Ca²⁺-dependent upstroke; fastest normal intrinsic rateInitiate the normal heartbeatNormal sinus P waves and regular rhythm
AV nodal cellsInteratrial septum near the tricuspid valveSlow conduction and intrinsic automaticity; important conduction delayLimit atrial-to-ventricular conduction and provide an escape pacemakerForms much of the PR interval; blocks can prolong or drop QRS complexes
Bundle of His and bundle-branch cellsInterventricular septumRapid conduction; fast-response action potentialsConduct impulses toward the ventriclesBundle-branch disease can widen or deform the QRS
Purkinje cellsEndocardial ventricular networkVery rapid conduction; automaticity at a slow escape rateActivate ventricular myocardium synchronouslySupport narrow, coordinated QRS depolarisation
Atrial working myocytesAtrial muscleFast-response action potential; contract when depolarisedAtrial contraction (“atrial kick”)Generate the P wave
Ventricular working myocytesVentricular myocardiumFast Na⁺ upstroke, plateau and prolonged repolarisationVentricular contraction and ejectionGenerate the QRS complex and T wave
Transitional / internodal cellsAtrial conduction pathwaysSpecialised atrial conduction; less automaticity than nodal tissueCarry SA-node impulses to the AV node and atriaContribute to normal atrial activation timing
Automaticity is the ability of a cell to depolarise spontaneously toward threshold without an external stimulus. In healthy adults the SA node normally dominates because it reaches threshold fastest.
Excitability is the ability to respond to a stimulus by generating an action potential. Conductivity is the ability to propagate that response through gap junctions to neighbouring cells.

Pacemaker hierarchy and intrinsic escape rates

Primary

SA node

Typical intrinsic rate: 60–100 beats/min in adults.

Secondary

AV junction

Typical intrinsic escape rate: 40–60 beats/min.

Tertiary

Ventricular Purkinje system

Typical intrinsic escape rate: 20–40 beats/min.

Rates vary with age, autonomic tone, fitness, illness and medication. Intrinsic rates are approximate and are used to understand escape rhythms.

Electrical pathway

Heart electrical conduction system and ECG waves

A normal impulse starts at the SA node, spreads through the atria, pauses at the AV node, then travels rapidly through the His–Purkinje network so that the ventricles depolarise in an efficient, coordinated sequence.

SA nodefires first · P wave begins AV nodedelay · PR interval Purkinje fibresrapid ventricular activation · QRS PathwaySA → atria → AV→ His → branches PurposeAtrial filling, thensynchronous pumping
Animation: impulse propagation through the atria, AV delay, and His–Purkinje activation. It is schematic rather than anatomically exact.

Normal conduction sequence

  1. SA node discharge initiates atrial depolarisation.
  2. Atrial activation spreads through the atria; contraction follows electrical activation.
  3. AV nodal delay permits ventricular filling before ventricular systole.
  4. His bundle and bundle branches conduct the impulse into the septum.
  5. Purkinje fibres rapidly activate the endocardium.
  6. Ventricular muscle depolarises from endocardium toward epicardium and contracts.
  7. Ventricular repolarisation restores the resting state and produces the T wave.

Representative conduction velocities

TissueApprox. velocityReason it matters
Atrial muscle0.3–1.0 m/sSpreads the P wave
AV node0.02–0.05 m/sCreates useful PR delay
His–Purkinje1.5–4.0 m/sEnables rapid, synchronised QRS
Ventricular muscle0.3–1.0 m/sCell-to-cell ventricular activation
Memory aid: slow at the AV node is normal; slow or blocked conduction elsewhere is often abnormal.

Annotated PQRST waveform

P PR QRS ST T R–R interval 0.04 s / small square (at 25 mm/s)Standard calibration: 10 mm = 1 mV
Click a segment button below to explore what each part of the tracing represents.
P wave: mainly atrial depolarisation, normally smooth and upright in most leads. It is usually less than 0.12 s wide.

Historical development

History of electrocardiography

The ECG emerged from discoveries in electrophysiology, precision instruments and clinical cardiology. Its history is an excellent case study in how sensors, amplifiers and signal processing transform physiology into a practical diagnostic technology.

1842Carlo Matteucci and the language of bioelectricityMatteucci demonstrated electrical activity in excitable tissue and helped establish the concept of the action current.
1869Early human cardiac recordingAlexander Muirhead is credited with an early recording of the heart’s electrical rhythm using a siphon recorder; the signal was difficult to interpret clinically.
1873Gabriel Lippmann’s capillary electrometerThe mercury-capillary device converted small voltage changes into visible movement, enabling early cardiac recordings.
1887Augustus D. Waller records a human ECGWaller used a capillary electrometer and demonstrated surface cardiac signals in a human subject, although the system was slow, distorted and not yet a practical clinical instrument.
1895Einthoven’s corrected curve and PQRST labelsWillem Einthoven mathematically corrected capillary-electrometer distortions and introduced labels that evolved into the P, QRS, T nomenclature still used today.
1903–1905String galvanometer and clinical ECGEinthoven developed a sensitive string galvanometer and achieved practical clinical recordings. On 22 March 1905, a successful human clinical ECG was recorded at Leiden.
1924Nobel PrizeEinthoven received the Nobel Prize in Physiology or Medicine “for his discovery of the mechanism of the electrocardiogram.”
1930s–1940sPrecordial leads and vector conceptsFrank Wilson and other investigators expanded surface recording to chest leads and clarified the idea that each lead views the heart from a different direction.
1950s–1970sMonitoring and ambulatory ECGPortable monitors, intensive-care systems and Norman Holter’s ambulatory recorder allowed prolonged rhythm assessment outside hospitals.
1980s–presentDigital and wearable electrocardiographyMicroprocessors, high-resolution ADCs, digital filters, wireless telemetry, smartphones and wearable devices have made ECG acquisition smaller, faster and more accessible.

Engineering lesson 1: sensitivity without fidelity is insufficient

Early devices could detect a cardiac signal but distorted its timing and shape. Einthoven’s key contribution combined a sensitive transducer with careful mathematical and instrumental correction.

Engineering lesson 2: clinical utility drives design

The modern ECG is not merely a waveform. It is a calibrated measurement system with standard paper speed, gain, lead geometry, filters, patient safety constraints and diagnostic reporting conventions.

Terminology: ECG is common globally; EKG is frequently used in the United States. Both refer to electrocardiography. “Electrocardiogram” is the recorded tracing.

The signal

The electrocardiogram (ECG)

An ECG records the time-dependent voltage difference between electrodes on the body surface. It is usually displayed in millivolts against time, with a standard diagnostic paper speed of 25 mm/s and calibration of 10 mm/mV.

Amplitude

Millivolt-scale

Typical surface signals are around 0.1–2 mV, so the acquisition system must amplify very small differential voltages.

Frequency

Approximately 0.05–150 Hz

Diagnostic ECG bandwidth commonly includes very low-frequency ST information and rapid QRS transitions. Device settings can differ.

Standard timing

25 mm/s

One small square is 0.04 s; one large square is 0.20 s. At 50 mm/s, these durations are halved.

What the standard components mean

ComponentPhysiological eventTypical adult referenceCommon clinical significance
P waveAtrial depolarisation<0.12 s; usually upright in I and IIAbsent, abnormal or fluttering P waves suggest rhythm or atrial disorders
PR intervalAtrial depolarisation plus AV nodal/His conduction0.12–0.20 sProlongation suggests AV block; short PR may suggest pre-excitation
QRS complexVentricular depolarisationUsually 0.08–0.10 s or lessWidening may indicate bundle-branch block, ventricular rhythm or conduction disease
ST segmentInterval between ventricular depolarisation and repolarisationNear baselineDepression may suggest ischaemia; elevation may suggest injury or infarction
T waveVentricular repolarisationUsually upright in many leads and concordant with QRSInversion or peaking may indicate ischaemia, strain or electrolyte disturbance
QT intervalTotal ventricular depolarisation and repolarisationRate-dependent; corrected QT is interpreted clinicallyProlongation increases arrhythmic risk; shortening has metabolic/genetic causes
U waveLate repolarisation-related activitySmall; not always visibleProminent U waves can be associated with hypokalaemia or bradycardia
Important: reference intervals vary with age, sex, lead, heart rate and measurement convention. Interpretation must be clinical, not based on one number alone.

Electrical vectors and why lead shape changes

A wave of depolarisation can be represented by a mean electrical vector. A lead measures the projection of that vector onto the lead axis. When the projection points toward the positive electrode, the deflection is mainly positive; when it points away, the deflection is mainly negative.

Lead axis + Depolarisation vector Projection Lead I conceptVector toward LA gives a positive deflectionVector away from LA gives a negative deflection Positive-dominant QRS Negative-dominant QRS
Original vector diagram: the surface ECG deflection depends on the relationship between the cardiac vector and the selected lead axis.

Automaticity

Pacemaker ability of the heart

The healthy heart can generate its own rhythm because SA-node cells possess automaticity. Their membrane potential does not remain constant between beats; it slowly drifts toward threshold during diastole.

Why the SA node normally wins

Pacemaker cells with the fastest phase-4 depolarisation reach threshold first and reset slower latent pacemakers. Sympathetic stimulation generally increases the slope and raises heart rate; parasympathetic activity generally slows it.

When latent pacemakers emerge

If SA-node impulses are too slow, blocked or absent, AV junctional or ventricular escape rhythms can take over. An escape rhythm is a safety response, not the primary healthy pacemaker.

Animated action-potential comparison

Phase
Phase 4
Elapsed
0 ms
Teaching mode
Contractile vs nodal
Guide to the phases: phase 4 is the resting or slow pacemaker potential; phase 0 is the rapid depolarisation upstroke; phase 1 is early repolarisation; phase 2 is the Ca²⁺-supported plateau; phase 3 is rapid K⁺-mediated repolarisation. The moving markers above show how the two action-potential shapes progress through these stages.

Mechanism of nodal automaticity

  1. After repolarisation, the nodal membrane begins a slow inward “funny” current, If.
  2. Outward K⁺ current decreases while inward Ca²⁺ currents increase.
  3. Threshold is reached and L-type Ca²⁺ channels drive the upstroke.
  4. K⁺ efflux repolarises the cell, and the cycle restarts.
Artificial pacemaker link: an electronic pacemaker replaces or supplements this biological timing when the native impulse is absent, too slow or intermittently blocked. A sensing circuit detects intrinsic activity, while a pulse generator delivers calibrated stimuli.

Clinical applications

Clinical conditions requiring an ECG

An ECG is requested when clinicians need evidence about heart rhythm, conduction, blood supply, chamber stress, electrolyte effects or device behaviour. Symptoms and context determine urgency.

Urgent or emergency presentations

  • Chest pain or pressure suggestive of myocardial ischaemia or infarction
  • Fainting, near-fainting or profound dizziness
  • Sustained palpitations with weakness, breathlessness or shock
  • Cardiac arrest, resuscitation or post-resuscitation care
  • Signs of severe electrolyte disturbance or drug toxicity affecting the heart

Routine or diagnostic indications

  • Known or suspected arrhythmia
  • Monitoring heart rate and rhythm during surgery or critical care
  • Hypertension, valvular disease or suspected chamber enlargement
  • Assessment before certain medications that affect QT interval
  • Pacemaker or ICD checks
  • Evaluation of breathlessness, fatigue, stroke or peripheral embolism

Representative ECG patterns and associated conditions

ECG observationPossible clinical associationEngineering/clinical caution
No normal P wave with irregularly irregular QRS spacingAtrial fibrillationRate may be irregular; automated counting can be unreliable
ST-segment elevation in contiguous leadsAcute myocardial injury or infarction, pericarditis or other causesMust be interpreted with symptoms, prior tracing and troponin testing
ST depression or T-wave inversionMyocardial ischaemia, strain, non-cardiac causesLead group and clinical context matter
PR interval >0.20 sFirst-degree AV blockConfirm calibration and measurement points
Wide QRS with bundle-branch morphologyConduction block or ventricular origin rhythmDistinguish conduction disease from artefact and lead placement error
Peaked T wavesHyperkalaemia or other causesConfirm with blood chemistry and clinical assessment
Prominent U waves or flattened T wavesHypokalaemia or drug effectsCompare with medication and electrolyte results
Low-voltage QRS complexesPericardial effusion, obesity, COPD, hypothyroidism or technical causesCheck electrode contact, gain and patient factors
Red-flag symptoms: severe chest pain, fainting, severe shortness of breath, cold sweat, weakness or palpitations with shock require urgent medical assessment. This guide is educational and is not a diagnostic tool.

Views of the heart

ECG leads

A standard 12-lead ECG uses 10 electrodes to create 12 electrical views: three bipolar limb leads, three augmented unipolar limb leads and six precordial chest leads. A lead is a measured voltage difference, not simply one wire.

RALALL Lead I: LA − RA Lead II Lead III Central terminalEinthoven relationship: II = I + III
Einthoven triangle: the standard limb leads lie approximately in the frontal plane.

Limb lead definitions

I = VLA − VRA
II = VLL − VRA
III = VLL − VLA
aVR = VRA − ½(VLA + VLL)
aVL = VLA − ½(VRA + VLL)
aVF = VLL − ½(VRA + VLA)
Lead II often gives a clear upright P wave and is commonly selected as a rhythm strip. A normal QRS axis is usually between −30° and +90°, though axis categories vary by reference.

Interactive lead-placement viewer

RALARLLLV1V2V3V4V5V6
V1: fourth intercostal space at the right sternal border; mainly views the right ventricle and septum.

Precordial positions

  1. V1: right sternal edge, 4th intercostal space
  2. V2: left sternal edge, 4th intercostal space
  3. V4: left mid-clavicular line, 5th intercostal space
  4. V3: midway between V2 and V4
  5. V5: anterior axillary line, level with V4
  6. V6: mid-axillary line, level with V4/V5

Lead groups used for localisation

Heart regionLeadsTeaching note
Inferior wallII, III, aVFST changes here may indicate inferior myocardial involvement
Lateral wallI, aVL, V5, V6Useful for lateral ischaemia and left-sided forces
Anterior wallV3–V4Chest leads provide horizontal-plane views
Septal regionV1–V2Bundle-branch morphology is often assessed here
Placement errors matter: high, low, reversed or poorly attached electrodes can mimic disease. RL is generally a reference/right-leg electrode and is not used to form one of the standard 12 displayed leads.

Quantitative reading

Heart rates from the ECG

Heart rate is calculated from the time between beats, usually the R–R interval. In sinus rhythm the atrial rate and ventricular rate are normally equal; in some heart blocks they differ.

Regular rhythm

HR (beats/min) = 60 ÷ R–R interval (seconds)
HR = 1500 ÷ small squares between R waves
HR = 300 ÷ large squares between R waves

Example: R–R = 4 large squares → 300/4 = 75 bpm.

Irregular rhythm / 10-second strip

Estimated average HR = number of QRS complexes × 6

A standard full-width rhythm strip is commonly 10 s long. Count QRS complexes over the full strip and multiply by six.

Interactive heart-rate calculators

Regular R–R interval

Heart rate
75 bpm
Interpretation
Normal

10-second rhythm strip

Average rate
102 bpm
Interpretation
Tachycardia
Adult reference: sinus rhythm is often described as about 60–99 bpm. Rates below 60 bpm are called bradycardia and rates above 100 bpm tachycardia, but symptoms and context are crucial.

Other useful rate concepts

Instrumentation

ECG recording: from patient to displayed waveform

The recording system must extract a millivolt differential signal in the presence of much larger common-mode interference, electrode offset and motion. Good clinical ECGs begin with careful preparation and continue through calibrated analogue and digital processing.

Patient +electrodesskin interface Protection +bufferinghigh input Z Instrumentationamplifiergain + CMRR Filters0.05–150 Hznotch optional ADC + DSPstoragecalibration Display /report12 leads Differential biosignal chain: transduction → safety → amplification → conditioning → digitisation → interpretation
Simplified diagnostic ECG signal chain. Actual devices include additional safety, isolation, lead-selection and diagnostic algorithms.

Typical engineering requirements

Stage / specificationTypical design objectiveWhy it matters
Electrode–skin interfaceClean, abraded or prepared skin; conductive gel; low and stable contact impedanceReduces motion artefact and 50/60 Hz pickup
Input impedanceVery high, commonly tens to hundreds of MΩLimits attenuation from high electrode impedance mismatch
Common-mode rejectionHigh CMRR, commonly >100 dBSuppresses interference common to both input electrodes
Amplifier gainOn the order of 500–1000 for a 1 mV ECGRaises millivolt signals to ADC range
Diagnostic bandwidthOften approximately 0.05–150 HzPreserves ST segments and rapid QRS detail
ADC resolutionCommonly 12–16 bits or moreProvides adequate dynamic range and small-signal resolution
Sampling rateCommonly 250–1000 samples/s per lead or higherSupports interval measurement and digital analysis
Patient isolationIsolation barrier and defibrillation protectionProtects the patient and equipment

Values are representative educational targets; actual systems follow applicable IEC and manufacturer specifications.

Common ECG artefacts and mitigation

ArtefactTypical appearanceCausePractical reduction
Baseline wanderLow-frequency driftRespiration, poor contact, movementSkin preparation, secure electrodes, high-pass filtering
Power-line interferenceRegular 50/60 Hz rippleElectromagnetic coupling, ground loops, high loop areaShort paired leads, reduce loop area, shielding, right-leg drive, notch filter when appropriate
Muscle noiseJagged high-frequency baselineEMG from tense or shivering musclesRelax patient, warm environment, filtering
Motion artefactLarge irregular deflectionsCable pull, electrode displacement, poor adhesionSupport cables, reattach electrodes, repeat tracing
Lead reversalAbnormal axis or inverted P/QRS in unexpected leadsRA/LA or precordial misconnectionConfirm labels and placement; repeat if suspected
AC coupling or saturationFlattened peaks or shifted baselineExcess electrode offset or amplifier overloadRe-prep skin, check electrode compatibility and device settings

Recording procedure

  1. Confirm patient identity, explain the test and obtain cooperation.
  2. Position the patient comfortably, usually supine, and protect privacy.
  3. Expose wrists, ankles and chest; avoid placing limb electrodes over major muscle masses.
  4. Shave hair if needed, clean and gently abrade skin, and apply gel or adhesive electrodes.
  5. Locate the sternal angle and fourth intercostal space before placing V1 and V2; place V4–V6 relative to these landmarks.
  6. Check cable integrity, patient relaxation and absence of movement/tremor.
  7. Record a calibrated 10-second 12-lead tracing; inspect signal quality before the patient leaves.
  8. Label and document symptoms, medications and relevant clinical context; repeat poor-quality recordings.

Virtual ECG signal laboratory

Current HR
75 bpm
Raw signal
Top trace
Processed signal
Bottom trace
Experiment: increase mains amplitude, then disable the notch. Add motion-like noise by reducing contact quality. Increase wander and toggle baseline correction. Observe why filtering cannot replace good electrodes and patient preparation.

Membrane electrophysiology

Depolarisation and repolarisation of a heart cell

Cardiac electrical activity results from changes in membrane permeability to ions. The inside and outside of a resting contractile myocyte have different charge, creating a resting membrane potential of roughly −90 mV.

Cardiac cell membrane Na⁺ outsideK⁺ insideCa²⁺ outside Depolarisation: inward Na⁺ / Ca²⁺Repolarisation: outward K⁺ Resting state: outside positive, inside negative (−90 mV)
Schematic ion movement during an action potential. Ion currents and membrane states are simplified.

Fast-response ventricular action potential

PhaseMembrane eventPrincipal ion movement
4Stable resting potentialHigh K⁺ permeability; inward rectifier K⁺ current helps maintain −90 mV
0Rapid depolarisation/upstrokeFast voltage-gated Na⁺ channels open; Na⁺ rushes inward
1Early repolarisation/notchNa⁺ channels inactivate; transient outward K⁺ current
2PlateauCa²⁺ influx is balanced by K⁺ efflux; contraction is sustained
3Rapid repolarisationCa²⁺ channels close; K⁺ leaves the cell
Contraction link: membrane depolarisation opens Ca²⁺ pathways. The resulting calcium transient activates the contractile machinery, converting the electrical event into mechanical work.

Wave propagation and recovery

Depolarisation makes the cell interior less negative and then positive. Charge transfer through gap junctions excites adjacent cells, producing a propagating wavefront. Repolarisation restores the negative internal potential and prepares the cell for another action potential.

Direction caution: a positive surface deflection is not simply “positive charge moving toward the electrode.” It depends on wavefront direction, spatial gradients, lead orientation and the distribution of active versus resting tissue.

Active learning

Practice, self-check and study plan

Mini-case studies

Case 1

Regular narrow-complex rhythm

R–R = 0.60 s, one P wave before each QRS, PR = 0.16 s.

Exercise: calculate rate, name the likely rhythm origin and list two checks before calling it normal.
Case 2

Noisy tracing

A 50 Hz ripple appears whenever a nearby mains-powered device is switched on.

Exercise: identify the interference source, propose three engineering remedies and explain any downside of an aggressive notch filter.
Case 3

Slow escape rhythm

No consistent P waves; regular wide QRS at about 35 bpm.

Exercise: explain why this may indicate a ventricular escape rhythm and why urgent clinical assessment is needed.

Multiple-choice self-test

Recommended study sequence

  1. Draw the conduction system and label the SA node, AV node, His bundle, bundle branches and Purkinje fibres.
  2. Memorise P, QRS and T wave events before memorising numerical intervals.
  3. Derive heart-rate equations from first principles: 25 mm/s → 1 small square = 0.04 s.
  4. Draw the acquisition block diagram and annotate each stage with one artefact it can introduce or reduce.
  5. Place the chest electrodes on a model or labelled diagram without looking.
  6. Use the simulations above to connect physiology, signal quality and clinical interpretation.

Revision

Summary and glossary

Must remember

  • P = atrial depolarisation
  • PR = atrial activity plus AV delay
  • QRS = ventricular depolarisation
  • ST = early ventricular recovery/plateau phase
  • T = ventricular repolarisation

Core equations

HR = 60 / RR(s)
HR = 300 / large squares
HR ≈ 6 × QRS in 10 s

Engineering priorities

  • Stable electrode contact
  • High input impedance
  • High CMRR
  • Appropriate gain and bandwidth
  • Safe isolation and calibrated ADC

Glossary

Action potentialVoltage change across a cell membrane caused by time-dependent ion currents.
ArtefactSignal contamination not produced by the intended cardiac activity.
AutomaticitySpontaneous impulse generation in pacemaker cells.
AV nodeAtrial conduction structure that slows impulse transmission to the ventricles.
Bipolar leadVoltage measured between two designated electrodes.
BradycardiaHeart rate below 60 bpm in adults.
CMRRCommon-mode rejection ratio; ability to reject identical interference on both inputs.
DepolarisationReduction of the negative resting membrane potential toward and beyond zero.
ElectrodeInterface that converts ionic current in tissue to electron current in a circuit.
Escape rhythmBackup rhythm from a subsidiary pacemaker when the primary impulse fails.
LeadConfigured electrical view of the heart, calculated from electrode voltages.
RepolarisationRestoration of the negative resting membrane potential after depolarisation.
Sinus rhythmRhythm initiated by the sinoatrial node.
TachycardiaHeart rate above 100 bpm in adults.
VectorDirection and magnitude representing a propagating cardiac electrical wavefront.

Selected further reading and sources

  1. Rivera-Ruiz M, Cajavilca C, Varon J. Einthoven’s String Galvanometer: The First Electrocardiograph. Texas Heart Institute Journal / PMC. https://pmc.ncbi.nlm.nih.gov/articles/PMC2435435/
  2. Hurst JW. Naming of the Waves in the ECG, With a Brief Account of Their Genesis. Circulation. https://www.ahajournals.org/doi/10.1161/01.cir.98.18.1937
  3. Khan SU et al. Introduction. I—Leads, rate, rhythm, and cardiac axis. PMC. https://pmc.ncbi.nlm.nih.gov/articles/PMC1122339/
  4. StatPearls / NCBI Bookshelf. Physiology, Sinoatrial Node. https://www.ncbi.nlm.nih.gov/sites/books/NBK459238/
  5. NCBI Bookshelf. Physiology, Cardiac Repolarization Dispersion and Reserve. https://www.ncbi.nlm.nih.gov/books/NBK537194/
  6. Nobel Prize. Willem Einthoven – Facts. https://www.nobelprize.org/prizes/medicine/1924/einthoven/facts/